INDICADORES BIOLÓGICOS
BT110
Rapid Self-contained Biological Indicator for Ethylene Oxide sterilization processes. Bacillus atrophaeus (ATCC 9372) 10^6 spores per vial. Readout: 4 hours.
An optional visual pH color change confirmation could be made after 48 hours of incubation. If sterilization process has not been successful, culture medium will change to a greenish color first, and then to yellow during incubation at 37 ºC, thus showing the presence of living spores. If sterilization process is successful culture medium will remain blue after the incubation process.
7-day readout for visual confirmation by color change is optional and not intended to be routinely performed; it is an initial validation of the 4 hour-reading. Fluorescence results may be compared to the 7-day visual reading.
NOTE: if 7-day readout is performed, a humidified environment will be required to avoid medium dry out.
FICHA TÉCNICA
PRODUCTO | BT110 |
CATEGORÍA | Indicador Biológico |
PROCESO | Óxido de etileno |
TIPO | 1 |
NORMA | ISO 11140-1 |
MARCA | Bionova® |
EMPAQUE | 50 unidades |
COLOR INICIAL | Magenta |
COLOR FINAL | Verde |
REACCIÓN | Univariable |
SOPORTE | Ampolleta |
MICRO ORGANISMOS | Bacillus atrophaeus (ATCC 9372) |
TIEMPO DE LECTURA | 4 horas |
REGLAMENTO | IRAM 37102-1, IRAM 37102-2, ISO 11138-1, ISO 11138-2 |
POBLACIÓN | 10^6 esporas/portador |